105
CANNON
®
Automatic Viscometer Models CAV-2100 and CAV-2200 with VISCPRO
®
Instruction & Operation Manual
Version 2g — May, 2009; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
NOTE
The VI Matcher does not match invalid samples UNLESS the Allow
Matching ... box from the Sample Filter tab options is checked. The VI
Matcher never matches unsuccessful samples (samples with no flow
time and a displayed kinematic viscosity of “0”). If one of the samples ran
successfully at one temperature but not at the other, the good sample
can still be used for VI matching if the failed sample is rerun within 8
hours.
VI Analysis filter
To adjust other VI analysis options, click the VI Report tab.
VI Report data filter options
Then use the check boxes to select which fields to include in the
report. After individual options have been selected, use the related
spin controls and radio buttons to determine the appearance of
data for kinematic viscosity, temperature and drop times.
8. When you have completed the configuration, click OK. The program
will prompt you to save the configuration.