Compare the values in the Tray Settings: Advanced window with your
archived values and make any necessary changes; then click OK.
Preparing the miniQV-X for testing
Inserting the miniQV-X temperature probes/reference thermometer
Inserting the probes
The temperature probe and thermal fuse probe are semi-permanently
installed. Initial installation or replacement requires removal of the top
cover of the bath unit. The probes are inserted through the narrow
aperture on the top bath flange and held in place with a sliding metal
plate (friction fitting) secured with a thumbscrew.
miniQV-X top bath flange
Inserting the thermometer
The ASTM reference thermometer/adapter assembly can be inserted
through the grommet in the bath top cover and through the larger circular
aperture on the top bath flange (see photo image above). The assembly
includes a stop collar with set screw that can be adjusted to vary the
thermometer height.
CAUTION
Do NOT insert a reference thermometer while the miniQV-X bath impel-
ler is operating, as damage to the thermometer and/or viscometer tube
may result. Make certain to insert the appropriate reference thermometer
for the current bath temperature! Use appropriate safety procedures
when handling the thermometer, as it contains mercury.